Part 6 · Chapter 19

Partial Discharge Measurement and Analysis

A withstand test asks whether insulation survives one deliberate overvoltage. But most insulation does not fail in a moment — it dies slowly, eroded from within by tiny sparks in a void no larger than a pinhead, each one harmless on its own yet relentless over years. This chapter is about catching that quiet decay: what a partial discharge is, why a cavity discharges while the bulk insulation holds, and how a few picocoulombs of charge betray a defect long before it becomes a failure.

High-Voltage Engineering Prof. Mithun Mondal Reading time ≈ 45 min
i What you'll learn
  • What a partial discharge is, and the three forms — internal voids, surface discharge and corona.
  • Why the gas in a cavity sees a field \(\varepsilon_r\) times higher than the solid and so discharges first.
  • The abc three-capacitor equivalent circuit and the inception voltage.
  • The apparent charge \(q = C_b\,\Delta V_c\), measured in picocoulombs, and why it is "apparent".
  • The IEC 60270 detection circuit — coupling capacitor and measuring impedance — and its calibration by an injected charge \(q_0 = C_0 V_0\).
  • Phase-resolved PD patterns that fingerprint defects, with inception (PDIV) and extinction (PDEV) voltages, and the acoustic, optical, chemical and UHF methods.
Section 19-1

What Is a Partial Discharge?

A partial discharge (PD) is a localised electrical discharge that bridges only part of the insulation between two conductors. It is not a full breakdown — the main insulation still holds and the apparatus keeps working — but a small spark has occurred at a weak spot, and it will occur again and again, every cycle, as long as the voltage is applied. Three forms dominate. An internal discharge happens inside a gas-filled void or cavity trapped in solid insulation. A surface discharge creeps along an interface where the tangential field is high, such as the edge of an electrode on an insulator. And corona is the discharge in gas at a sharp point or thin wire, met already in the breakdown chapters.

What makes PD so important is that it is both a symptom and a cause. It is a symptom because a measurable discharge betrays a defect — a void left in casting, a delamination, a sharp burr — that should not be there. It is a cause because each tiny spark bombards the cavity wall, eroding the material, growing conductive channels called electrical trees, and over months or years carving a path that finally becomes a full breakdown. Detecting PD therefore predicts failure long before it happens, which is why a PD test is among the most valued routine and diagnostic tests of all.

internal (void) surface corona (point)
Three forms of partial discharge — an internal void in solid insulation, a surface discharge along an interface, and corona at a sharp point in gas
Section 19-2

Field Enhancement in a Cavity

Why should a small gas void discharge while the solid around it, far stronger, stays intact? The answer is a piece of field theory from the very first chapters. At the boundary between the solid and the gas, the normal component of the electric flux density \(D = \varepsilon E\) is continuous. The gas has permittivity \(\varepsilon_0\) and the solid \(\varepsilon_0\varepsilon_r\); for a flat void lying across the field, continuity of \(D\) forces the field inside the gas to be \(\varepsilon_r\) times larger than in the surrounding solid:

Field in a flat void
\[ \varepsilon_0\,E_{\text{cav}} = \varepsilon_0\varepsilon_r\,E_{\text{diel}} \qquad\Longrightarrow\qquad E_{\text{cav}} = \varepsilon_r\,E_{\text{diel}} \]
+V gas void E in solid E_cav = ε_r·E
The void sees a stronger field — continuity of \(D\) across the boundary makes \(E_{\text{cav}}=\varepsilon_r E_{\text{diel}}\), while the gas also breaks down at a far lower strength than the solid

So the void suffers a double disadvantage. It experiences a field several times stronger than the bulk (a typical \(\varepsilon_r\) of 3–5), and the gas inside it breaks down at a far lower strength than the solid that surrounds it — air gives way at about \(3~\mathrm{kV/mm}\) while good solid insulation withstands tens of kilovolts per millimetre. The combination guarantees that, as the applied voltage rises, the gas in the cavity is the first thing to break down, sparking over while every other part of the insulation comfortably holds. That first spark is the partial discharge.

Section 19-3

The abc Model and Inception

To turn that picture into something measurable, the cavity is represented by a three-capacitor equivalent circuit — the classic abc model. The cavity itself is a capacitor \(C_c\); the slice of healthy dielectric directly in series with it (above and below the void) is \(C_b\); and all the remaining sound insulation in parallel is \(C_a\). Before any discharge, the applied voltage divides capacitively, placing across the cavity the fraction

Cavity voltage
\[ V_c = V\,\frac{C_b}{C_b + C_c} \]
V C_a (bulk) C_b C_c (void) fires at V_i
The abc model — \(C_a\) is the bulk dielectric, \(C_b\) the slice in series with the void, and \(C_c\) the cavity, which discharges like a tiny spark gap once its voltage reaches the inception level

The cavity \(C_c\) behaves like a miniature spark gap. As the AC voltage climbs, \(V_c\) climbs with it until it reaches the gas breakdown value — the cavity's own little inception level. At that instant the void sparks over, its voltage collapses almost to zero, and the discharge is done. The terminal voltage at which this first occurs is the partial discharge inception voltage, and below it the insulation is silent. Each time the applied wave drives \(V_c\) back up to inception, the cavity fires again, producing the characteristic train of pulses, several per half-cycle, clustered on the rising flanks of the voltage.

Section 19-4

Apparent Charge — the Measurable Quantity

When the cavity fires, a real charge moves inside the void — but that charge is buried in the insulation and cannot be reached. What the external circuit can sense is the sudden small change it forces at the terminals: the collapse of the cavity voltage by \(\Delta V_c\) pushes a pulse of charge through \(C_b\) into the measuring circuit. This terminal-measurable quantity is the apparent charge \(q\):

Apparent charge
\[ q = C_b\,\Delta V_c \qquad (\text{measured in picocoulombs, pC}) \]

It is called "apparent" because it is not the true charge that flowed in the void — that real charge, involving \(C_c\) as well, is always larger and is inaccessible. The apparent charge is simply the charge that, if injected at the terminals, would produce the same terminal effect as the discharge. That is exactly the quantity we can calibrate and read, so PD severity is universally quoted as an apparent charge in picocoulombs: a few pC in good insulation, hundreds or thousands of pC in deteriorating apparatus. A rising apparent charge over time is a clear warning that a defect is growing.

We measure the shadow, not the spark. The real discharge inside the void is hidden; the apparent charge is its measurable shadow at the terminals. It under-states the true charge but is reproducible and calibratable — which is exactly what a diagnostic number must be.
Section 19-5

The Detection Circuit

A PD pulse is a tiny, very fast event — a few picocoulombs in a few nanoseconds — riding on top of a large power-frequency voltage. Pulling it out is the job of the straight detection circuit standardised in IEC 60270. The test object \(C_x\) is energised, and beside it a coupling capacitor \(C_k\) is connected in parallel. When the cavity fires, the fast pulse needs a low-impedance path at high frequency, and \(C_k\) provides it: the pulse circulates between \(C_x\) and \(C_k\), passing through a measuring impedance \(Z_m\) placed in series with one of them. Across \(Z_m\) the pulse develops a small voltage that is amplified, filtered and displayed, while the power-frequency voltage — being low-frequency — is blocked from the detector.

HV~ Z (block) C_x (object) C_k (coupling) Z_m amp + display
The IEC 60270 straight detection circuit — the coupling capacitor \(C_k\) gives the fast PD pulse a path, and the measuring impedance \(Z_m\) turns it into a signal for the amplifier, while the power-frequency voltage is blocked

Two refinements matter in practice. The detector may be narrow-band (tuned to a chosen frequency for good noise rejection) or wide-band (preserving the pulse shape for analysis). And because the signals are so small, external interference — radio, switching noise, corona on the leads — easily swamps a real PD; a balanced (bridge) circuit that compares the test object against a discharge-free reference can cancel common interference, letting the genuine internal discharge stand out.

Section 19-6

Calibration and the pC Scale

The detector produces a voltage pulse, but we want an answer in picocoulombs, and the relationship between them depends on the stray capacitances of the whole set-up. The cure is direct calibration: before the test, a known charge is injected at the terminals of the test object and the detector's response is noted, fixing the scale. The known charge is produced by stepping a small, accurately known voltage \(V_0\) across a small, accurately known capacitor \(C_0\):

Calibration charge
\[ q_0 = C_0\,V_0 \]

A calibrator that switches, say, \(5~\mathrm{V}\) across a \(1~\mathrm{pF}\) capacitor injects a clean \(5~\mathrm{pC}\) pulse; the detector deflection it produces then defines "\(5~\mathrm{pC}\)" on the screen, and every later reading is scaled against it. Calibration must be done on the actual circuit, with the actual test object connected, because it is the capacitances of that arrangement that set the response — a calibration is not transferable from one object to another. With it done, the instrument reports apparent charge directly, and the diagnostician can compare today's reading against the standard's permitted level and against the object's own history.

Section 19-7

Phase-Resolved Patterns and Defect Fingerprints

The magnitude of the apparent charge tells you how much discharge there is; the pattern tells you what kind. Because PD pulses fire at preferred parts of the AC cycle, plotting each pulse's magnitude against the phase angle of the voltage builds a picture — the phase-resolved PD (PRPD) pattern — whose shape is a fingerprint of the defect. Internal voids tend to discharge symmetrically on the rising flanks of both half-cycles, giving twin rabbit-ear clusters; surface discharges skew differently between the half-cycles; and corona at a point appears strongly in one polarity, betraying its asymmetry.

phase → q + half – half symmetric → internal void
A phase-resolved PD pattern — pulse clusters plotted against the cycle phase; symmetric clusters on the rising flanks of both half-cycles point to an internal void

Two voltages complete the diagnostic. The PD inception voltage (PDIV) is the level at which discharges first appear as the voltage is raised; the PD extinction voltage (PDEV) is the level at which they cease as it is lowered. The extinction voltage is always the lower of the two, because once the cavity has been ionised it re-discharges more easily — residual charge and free electrons lower the bar — so the voltage must drop below inception before the discharges stop. A healthy margin between the working voltage and the PDIV is a primary design goal, and a PDIV that falls over time is a clear sign of progressing damage.

Section 19-8

Other Methods and Why PD Matters

The electrical method just described is the reference, but a discharge announces itself in several ways at once, and each has its own detector. A PD makes a faint sound — an ultrasonic click — that acoustic sensors can pick up and even locate by timing. It emits light, so an optical method using ultraviolet-sensitive cameras can see surface discharge and corona in the open. It produces chemical by-products, so in an oil-filled transformer dissolved-gas analysis (DGA) reads the gases the discharge cooks out of the oil, distinguishing PD from overheating and arcing. And it radiates radio-frequency energy, so the UHF method, sensing emissions in the hundreds of megahertz, has become the standard for gas-insulated switchgear where the metal enclosure both shields external noise and guides the UHF signal to internal antennas.

MethodSensesStrengthTypical use
Electrical (IEC 60270)apparent charge (pC)quantitative, calibratablefactory & lab reference
Acoustic / ultrasonicsound of the sparklocates the sourcetransformers, on-site
Optical / UVemitted lightsees corona & surface PDoutdoor lines, insulators
Chemical (DGA)dissolved gases in oildistinguishes PD/arc/heatoil-filled transformers
UHFradio emission (~MHz–GHz)noise-immune, sensitivegas-insulated switchgear

Taken together these methods make PD the foundation of modern insulation diagnostics. A withstand test (Chapter 18) gives a single pass/fail at a moment in time; a PD measurement reveals the slow internal health of the insulation and how it changes — which is why it sits at the heart of the condition-monitoring strategies of Chapter 21, alongside the dielectric-loss measurement we turn to next.

Section 19-9

Worked Examples

1 Will the void discharge?

Problem. Solid insulation of \(\varepsilon_r = 4\) carries a field of \(2~\mathrm{kV/mm}\). A flat gas void lies across the field. Find the field in the void and decide whether it discharges (air breaks down at about \(3~\mathrm{kV/mm}\)).

Solution. Use \(E_{\text{cav}} = \varepsilon_r E_{\text{diel}}\):

Working
\[ E_{\text{cav}} = 4\times2 = 8~\mathrm{kV/mm} \;\; > \;\; 3~\mathrm{kV/mm} \quad\Rightarrow\quad \text{discharges} \]

The void sees 8 kV/mm, well past air's strength, so it sparks while the solid — needing far more — holds. This is why a harmless-looking field in the bulk still produces PD.

2 Cavity voltage from the abc model

Problem. In an abc model, \(C_b = 8~\mathrm{pF}\) and \(C_c = 2~\mathrm{pF}\). The voltage across the series branch is \(10~\mathrm{kV}\). Find the cavity voltage.

Solution. Apply \(V_c = V\,C_b/(C_b+C_c)\):

Working
\[ V_c = 10\times\frac{8}{8+2} = 10\times0.8 = 8~\mathrm{kV} \]

Most of the branch voltage falls across the small cavity capacitance, so the void carries 8 kV — the reason a thin void in a thick dielectric reaches its breakdown so readily.

3 Apparent charge

Problem. When the void fires, its voltage collapses by \(\Delta V_c = 50~\mathrm{V}\). With \(C_b = 8~\mathrm{pF}\), find the apparent charge.

Solution. Apply \(q = C_b\,\Delta V_c\):

Working
\[ q = (8\times10^{-12})(50) = 4\times10^{-10}~\mathrm{C} = 400~\mathrm{pC} \]

The measurable apparent charge is 400 pC — a clearly detectable level that would flag this void for attention. The true void charge is larger, but inaccessible.

4 Sizing the calibrator

Problem. A calibrator must inject a \(10~\mathrm{pC}\) reference pulse using a \(2~\mathrm{pF}\) capacitor. What step voltage is needed?

Solution. From \(q_0 = C_0 V_0\), solve for \(V_0\):

Working
\[ V_0 = \frac{q_0}{C_0} = \frac{10\times10^{-12}}{2\times10^{-12}} = 5~\mathrm{V} \]

A clean 5 V step across the \(2~\mathrm{pF}\) capacitor delivers the \(10~\mathrm{pC}\) calibration pulse that fixes the picocoulomb scale.

5 Inception and extinction

Problem. An object shows PD starting at \(12~\mathrm{kV}\) as the voltage is raised. As the voltage is lowered the discharges persist down to \(9~\mathrm{kV}\). Identify the PDIV and PDEV and explain the difference.

Solution. The higher level is inception, the lower is extinction:

Identification
\[ \text{PDIV} = 12~\mathrm{kV}, \qquad \text{PDEV} = 9~\mathrm{kV}, \qquad \text{PDEV} < \text{PDIV} \]

PD ceases only at 9 kV, below the 12 kV at which it began, because the ionised void re-fires more easily once started — so the working voltage should sit safely below the 12 kV inception level.

Review

Chapter Summary

What PD is

A localised spark that bridges only part of the insulation — internal void, surface or corona. Both a symptom of a defect and a cause of slow failure.

Why the void fires

Continuity of \(D\) gives \(E_{\text{cav}}=\varepsilon_r E_{\text{diel}}\), and gas is weaker than solid, so the cavity discharges first.

abc model

\(C_a\), \(C_b\), \(C_c\); the cavity sees \(V_c = V\,C_b/(C_b{+}C_c)\) and fires at the inception level like a tiny spark gap.

Apparent charge

\(q = C_b\,\Delta V_c\) in pC — the measurable "shadow" of the hidden real charge, the universal severity figure.

Detect & calibrate

IEC 60270: coupling capacitor \(C_k\) + measuring impedance \(Z_m\); calibrate by injecting \(q_0 = C_0 V_0\) on the actual circuit.

Patterns & methods

Phase-resolved patterns fingerprint defects; PDEV < PDIV. Acoustic, optical, DGA and UHF complete the diagnostic toolkit.

Practice

Problems

For each item, first identify what it tests — the nature of PD, the field-enhancement reason, the abc model, the apparent charge, the detection/calibration circuit, or the patterns — then apply it. Difficulty rises down the list.

  1. Define a partial discharge and explain why it is described as both a symptom and a cause of insulation failure.
  2. Name the three forms of PD and give a physical example of each.
  3. A flat void sits in insulation of \(\varepsilon_r = 3.5\) where the bulk field is \(2.5~\mathrm{kV/mm}\). Find the void field and judge whether it discharges (air \(\approx 3~\mathrm{kV/mm}\)).
  4. Draw and label the abc equivalent circuit, stating what each capacitor represents.
  5. For \(C_b = 6~\mathrm{pF}\), \(C_c = 3~\mathrm{pF}\) and a branch voltage of \(9~\mathrm{kV}\), find the cavity voltage.
  6. Explain why the apparent charge is "apparent", and why it is nonetheless the quantity that is measured and quoted.
  7. A void discharge drops the cavity voltage by \(40~\mathrm{V}\) with \(C_b = 10~\mathrm{pF}\). Find the apparent charge in pC.
  8. In the detection circuit, state the role of the coupling capacitor and of the measuring impedance, and why the power-frequency voltage does not appear at the detector.
  9. A calibrator injects \(20~\mathrm{pC}\) using a \(4~\mathrm{pF}\) capacitor. Find the step voltage, and explain why calibration must be done on the actual test circuit.
  10. Explain why the PD extinction voltage is lower than the inception voltage, and why a phase-resolved pattern can distinguish an internal void from corona.
Tip: this chapter follows one chain of cause and effect. A weaker, more-stressed gas void (\(E_{\text{cav}}=\varepsilon_r E_{\text{diel}}\)) fires first; the abc model turns that into an inception voltage and a charge collapse; the collapse pushes an apparent charge \(q=C_b\Delta V_c\) into a detection circuit calibrated by \(q_0=C_0V_0\); and the timing of the pulses, read as a phase pattern, names the defect. Master that chain and every formula in the chapter is a link in it.